Steps to change the Bioptech's Controller Fuse Holder from 110V to 220V or 220V to 110V

 

1) Operating voltage range is selected by the orientation of the fuse holder located in the Power Input Module on the back surface of the Controller. Confirm the operating voltage by checking the orientation of the fuse holder. Operating voltage is indicated by the alignment of the arrows (shown above). In this case the controller is set to 110-120V.

 

2) If the operating voltage needs changed, the fuse holder can be removed by carefully prying it out of its receptacle as shown with a flat edge screwdriver.

 

3) When reinstalling the fuse holder make sure the arrows align to the proper voltage. For 90-130 supply line voltage place the fuse block with the arrow pointing to 110-120V. For 180-260 supply line voltage place the fuse block with the arrow pointing to 220-240V.

 

 