|
|
|
Objective Heater, Standard |
16mm – 28mm |
Temperature Range |
Ambient to 430C 6 Watts |
|
Objective Heater, Large |
26mm – 35mm |
Temperature Range |
Ambient to 430C 6 Watts |
|
Sensor |
Thermistor |
Temperature Stability |
± 0.10C Typically |
PLEASE
READ BEFORE USING THE BIOPTECHS, INC. OBJECTIVE HEATER!
Objectives are expensive and delicate.
We recommend following the directions below carefully.
Although great care has been taken to produce a reliable product
to control the temperature of your objective, there is always some risk.
BIOPTECHS, INC. assumes no responsibility for use of this or
any other objective heater. The
user assumes all liability for any damage that may occur due to the use
of this objective heater.
The controller can operate between 90 -
260V 50-60Hz. The AC power
line input, power switch, voltage selector, and fuses are all
incorporated on the rear panel. The
fuse carrier contains two 0.5A slow blow fuses.
The fuses can be accessed by prying the fuse block out of its
socket with a small screwdriver wedged under the small slot in the
opening of the AC line receptacle.
When replacing the fuse block, make sure to align the arrow on
the outside of the fuse block with the closest supply line voltage:
§
For 90-130 supply line voltage place the fuse block with the arrow
pointing to 110-120V.
§
For 180-260 supply line voltage place the fuse block with the arrow
pointing to 220-240V.
Controller
Operation:
1.
Turn the unit on using the main on-off switch located on the rear
panel. All further
adjustments are made on the front panel.
DO NOT plug in the
objective heater at this time.
2.
Turn
the display switch to the "OBJECTIVE" position.
The temperature display will read a number between
3.
Set
the display switch to the "SET" position, then insert the
enclosed screwdriver into the hole marked "SETPOINT" and
adjust the controller 1.8° less than the desired setpoint to compensate
for the thermal characteristics of the objective.
4.
Turn
the controller OFF.
5.
Mount
the objective heaterband onto the objective.
(See illustration – assembled objective heater)
6.
Plug
the objective heater into the controller then turn the controller ON.
7.
IMMEDIATELY set the display knob to the objective position and
check that it is indicating a value between ambient and setpoint.
If the temperature value displayed is not realistic turn the
controller off immediately. (Other
values indicate possible damage to the temperature probe).
As a safety feature, when the sensor temperature reaches setpoint
minus 0.9°C an internal latch is set.
Then if the temperature of the sensor falls
below 0.9°C less than setpoint, an alarm
will sound and the heater current will be shut off. The controller will still be on but the output disabled.
If the red reset button located next to the red LED is pressed it
will return the controller to cold startup mode and current will return
to the heaterband.
8.
Always make sure the sensor is in direct contact with the
objective. When you press
the reset button, the controller always enters cold startup mode and
ignores out of range conditions. Therefore
it is important that the objective heater is installed properly.
9.
The
objective will come up to temperature in about 5 to 10 minutes.
The controller will maintain a constant temperature relative to
the value of the contact probe. However,
it may be necessary to fine-tune the setpoint adjustment for your
particular objective. The
temperature of the upper element of the objective should be monitored
with an independent thermal sensor to establish the correct setpoint
value.
Heater
Installation: (The heater band
should be unplugged from the controller and the controller shut OFF)
1.
With the objective secured in the nosepiece of the microscope
hold the heater assembly upright with the loop in a superior position. (See illustration – assembled objective
heater)
2.
Insert the enclosed 2mm Allen screwdriver into the hole in the
top of the metal enclosure. Rotate
the Allen screwdriver clockwise to allow the loop to expand to fit
various objectives. Please
note that it may be necessary to gently withdraw the black heater band
from the metal enclosure to relieve the slack in the tensioning
mechanism. DO
NOT place tension on the red wires.
3.
Place the heater loop over the exposed cylindrical end of the
objective closest to the specimen making sure the band is evenly applied
to this area and that the temperature sensor is also making direct
contact with the objective.
4.
Tighten the heater loop onto the objective by turning the Allen
screwdriver counterclockwise until the loop is snug.
Use only enough pressure to cause the heater to conform to the
objective diameter and uniformly compress the temperature sensor to the
objective. DO NOT OVER
TIGHTEN the loop as damage will occur to the tensioning mechanism.
5.
There is a slight mechanical offset to the temperature sensor to
maintain contact which may cause the tension mechanism to rise slightly. It may be necessary to restrict this movement with an elastic
band.
6.
Plug the Objective Heater band into the Controller then turn the
controller ON. (Refer to
step seven above)

Note: To reduce the possible detrimental effects of temperature cycling, it is recommended that the objective be kept in a warming box when not in use.
©Bioptechs,
Inc., 11/2/00
BNi04200302
|
[Home] [Products] [Price list] E-mail:info@bioptechs.com |