
Objective Heater System
Instructions
Before You Begin:
·
Check to make sure there is no
visible damage in shipment and save all packing materials.
·
The Controller box should
contain; controller, power cord, 2.5mm screwdriver, 2mm Allen driver,
1/8” jack, and 3 pin mini DIN connector.
·
Check the packing list and call
your supplier if there are any discrepancies.
The
Bioptechs Objective Temperature Control System is comprised of two
components. One part is the
electronic controller unit; the other is an adjustable objective heater
loop which attaches to high numeric aperture (immersion) microscope
objectives. The objective
heater is designed to eliminate the heat-sink effect that is accelerated
by the presence of an optical coupling medium; oil, glycerin or water.
These objectives have a retractable central tube which contains
the optical elements. Heat
is directly transferred to this tube and sensed by a thermistor which is
part of an intelligent feedback control loop.
The heater loop requires a minimum of 5mm longitudinal physical
contact with the cylindrical objective surface.
In some cases there may be a decorative collar or sleeve on the
objective which must be removed in order to permit adequate surface
contact for the heater loop. If
an outer collar or sleeve is present we recommend using Bioptechs
Objective Wrench and Pliers for proper removal (Bioptechs Part Number:
021523; See figure 3
– outer sleeve removed).
Directions
PLEASE
READ BEFORE USING THE BIOPTECHS, INC. OBJECTIVE HEATER!
Objectives
are expensive and delicate. We
recommend following the directions below carefully. Although great care has been taken to produce a reliable
product to control the temperature of your objective, there is always
some risk. BIOPTECHS,
INC. assumes no responsibility for use of this or any other objective
heater. The user assumes all liability for any damage that may occur
due to the use of this objective heater.
Front
Panel of Objective Heater Controller

Figure
1
Control
Interface Descriptions:
Numeric
Display should be illuminated when
the controller is turned on. If it does not light, check the main power source and the
fuse block located in the power input module where the AC power cord
plugs in.
Display
Selector is the black button
adjacent to the display. It
selects the source of the information displayed.
When the controller is turned on, the default condition is to
display the setpoint value indicated by the red light adjacent to the
word Setpoint.
The button selects one of a sequence of display options.
This button can be pressed at any time and has no effect on the
temperature regulation functions of the controller.
Display
Lights indicate the source of the
value displayed on the numeric display.
Setpoint
is adjusted by a potentiometer accessible with the 2.5mm screwdriver
provided. To adjust the
setpoint, set the numeric display to Setpoint
and adjust the potentiometer to the desired value.
If you need to change the setpoint value during your experiment
on a routine basis, there is a remote setpoint jack on the back of the
controller. See Remote
Setpoint.
Objective
is the location where the
calibration of the circuit reading the objective thermistor can be
adjusted.
Reference
is the location where the
calibration of the circuit reading the reference probe can be adjusted.
Reset
(Red Button) resets the controller if the alarm sounds.
Note: If the alarm
went off there must have been a reason.
Check the objective before resetting the controller.
Calibrate
substitutes precision resistors into the circuit that reads the
thermistors so that calibration adjustments can be made to a known
value.
Objective
connector is a six pin mini DIN connector where objective heater is
plugged in.
Back
Panel of the Objective Controller

Figure
2
AC Power
line input, power switch, voltage selector, and fuses are all
incorporated in the AC Power Input Module.
The controller can operate between 90 - 260V 50-60Hz, or a 24V
battery can be used as an alternate power supply.
The fuse carrier contains two 0.5A slow blow fuses.
The fuses can be accessed by prying the fuse block out of its
socket with a small screwdriver wedged under the small slot in the
opening of the AC line receptacle.
When replacing the fuse block, make sure to align the arrow on
the outside of the fuse block with the closest voltage value to your
supply line voltage:
·
For 90-130 supply line voltage
place the fuse block with the arrow pointing to 110-120V.
·
For 180-260 supply line voltage
place the fuse block with the arrow pointing to 220-240V.
Temp
Out is a voltage output direct from
two instrument amplifiers reading the thermistors from channel A and B.
The output from this
socket is temperature / 10. 37.0°C
= 3.70 Volt DC.
The left pin is channel
A, center pin is channel B, and right pin is ground.
Remote
Setpoint is an analog input that
allows the user to set the controllers setpoint from an external DC
source. The sleeve is earth ground. A DC voltage can be applied to the tip of this jack equal to
temperature /10. i.e.: 28.5°C
= 2.85 Volt. When the jack
is inserted into the controller it will automatically switch to the
jacks value. When removed
it will default to the setting on the front of the controller.
The remote setpoint can be used for programming or cycling
operating temperatures.
Controller
Operation:
-
Turn the unit on using the main on-off switch located on the rear
panel. All further
adjustments are made on the front panel.
DO NOT plug the
objective heater into the Objective Controller at this time.
-
Turn the display switch to the "OBJECTIVE" position.
The temperature display will read a number between -3.7 and -4.2.
This is normal. Press
the black button labeled "CALIBRATE" and make sure the display
reads 25.0°C. If
adjustment is necessary, turn the screw in the hole labeled
"OBJECTIVE" with the enclosed screwdriver.
CW = increase, CCW = decrease.
-
Set the display to the "SETPOINT", then insert the
enclosed screwdriver into the hole marked "SETPOINT" and
adjust the controller 1.8° less than the desired setpoint to compensate
for the thermal characteristics of the objective.
-
Turn the controller OFF.
-
Mount the objective heater band onto the objective.
(See figure 3 – assembled objective heater)
-
Plug the objective heater into the controller.
Then turn the
controller ON.
-
IMMEDIATELY set the display knob to the objective position and
check that it is indicating a value between ambient and setpoint.
If the temperature value displayed is not realistic, turn the
controller off immediately. (Other
values indicate possible damage to the temperature probe).
As a safety feature, when the sensor temperature reaches setpoint
minus 0.9°C an internal latch is set.
Then if the temperature of the sensor falls below 0.9°C less
than setpoint, an alarm will sound and the heater current will be shut
off. The controller will
still be on but the output disabled.
If the red reset button located next to the red LED is pressed it
will return the controller to cold startup mode and current will return
to the heater band.
-
Always make sure the sensor is in direct contact with the
objective. When you press
the reset button, the controller always enters cold startup mode and
ignores out of range conditions. Therefore, it is important that the objective heater is
installed properly.
-
The objective will come up to temperature in about 5 to 10
minutes. The controller
will maintain a constant temperature relative to the value of the
contact probe.
-
It may be necessary to fine-tune the setpoint adjustment for your
particular objective. The
temperature of the upper element of the objective can be monitored with
the thermistor sensor (small bead on the end of the fine wire extending
from the side of the objective heater assembly known as the reference
thermistor) to establish the correct setpoint value.
Heater Installation:
(The heater band should be unplugged from the controller and the
controller shut OFF)
-
With the objective secured in the nosepiece of the microscope
hold the heater assembly upright with the loop in a superior position. (See
figure 3 – assembled objective heater)
-
Insert the enclosed 5/64mm or 2mm Allen screwdriver into the hole in the
top of the metal enclosure. (See figure
4) Rotate
the Allen screwdriver clockwise to allow the loop to expand to fit
various objectives. Please note that it may be necessary to gently withdraw the
black heater band from the metal enclosure to relieve the slack in the
tensioning mechanism. DO NOT place tension on the heater band wires.
-
Place the heater loop over the exposed cylindrical end of the
objective closest to the specimen making sure the band is evenly applied
to this area and that the temperature sensor is also making direct
contact with the objective.
-
Tighten the heater loop onto the objective by turning the Allen
screwdriver counterclockwise until the loop is snug.
Use only enough pressure to cause the heater to conform to the
objective diameter and uniformly compress the temperature sensor to the
objective. DO NOT OVER
TIGHTEN the loop as damage will occur to the tensioning mechanism.
-
There is a slight mechanical offset to the temperature sensor to
maintain contact that may cause the tension mechanism to rise slightly.
It may be necessary to restrict this movement with an elastic
band.
-
Plug the Objective Heater band into the
controller. Then turn the
controller ON. (Refer to step seven above)
Note:
To reduce the possible detrimental effects of temperature
cycling, it is recommended that the objective be kept in a warming
box when not in use. We
recommend the Boekel Warmer (Bioptechs Part Number:
260700)
Click Here
to view a video of assembly.
Objective Heater.mpg
~10MB

*** CAUTION ***
Excessive use of
immersion oil on an objective is detrimental to both the objective and
the heater-band. Immersion oil should not be allowed to come into
contact with the heater-band. If contact occurs the heater-band can be
cleaned with a tissue dampened with alcohol.

Figure
3

Figure 4
Specifications
| Objective
Heater, Standard |
16mm
– 28mm |
Temperature Range |
Ambient
to 500C 6
Watts |
| Objective
Heater, Large |
26mm
– 35mm |
Temperature Range |
Ambient
to 500C 6
Watts |
| Sensor |
Thermistor |
Temperature Stability |
±
0.10C Typically |
|